• BACT/ALERT® 3D Dual-T

BACT/ALERT®3D Dual-T

Fully automated microbial detection system

  • Rapidity
  • Traceability
  • Ease-of-use
Hozzáad

További tájékoztatást szeretne?

The BacT/ALERT® 3D Dual-T apparatus was designed for users who need to perform rapid and easy product control. Performing non-destructive, dual-temperature microbial detection, this alternative system offers its users the most reliable results.

Automation Meets Tradition

The BacT/ALERT® Dual-T apparatus represents the automation of a traditional method : after injecting culture media bottles with a given sample, users only need to load the bottles into their system. Both the incubation and reading phases are automatically and objectively performed :

innovative method for microbial detection & sterility testing

    • Simplified testing process
    • Reduced “hands-on” time and labor costs

     

    Designed with our own patented colorimetric technology, this growth-based system can detect within 24-72 hours an extremely large range of organisms, such as aerobic, anaerobic, facultative microorganisms, yeast and fungi.

     

    The dual-temperature feature of this system guarantees a complete coverage of growth conditions – and is compliant with the different pharmacopoeias - allowing to conduct microbial detection at both 32.5°C and 22.5°C.

    To ensure accurate results’ traceability, we provide you with a 21 CFR Part 11-compliant data management system. Depending on your microbial testing needs, we offer the possibility to choose the most appropriate system among three different options : Signature, Select and Select Link (find out more in our e-catalogue)

    bioMérieux also offers a wide range of BacT-ALERT® 3D Dual-T culture media bottles, adapted to the type of microbial detection performed.

     

    Benefits of the BacT/ALERT® 3D Dual-T system

    • Reliable test results
    • Ease of use
    • Simplified workflow
    • Rapid detection
    • Non-destructive sterility test results
    • Flexible Data Management Options.

    BacT/ALERT® 3D Dual-T apparatus Implementation guide

    To help you switch from a traditional 14-day testing method to an alternative automated system, bioMérieux will provide with the BacT/ALERT® 3D Dual-T Implementation Guide (BIG), containing the following:

    - Installation Qualification (IQ) protocol

    - Operational Qualification (OQ) protocol

    - Performance Qualification (PQ) protocol

     

     

     

    System electrical requirements:

     

    100/120 Volts (50-60 Hz)

    220/240 Volts (50-60 Hz)

     

    Power consumed:

     

    At 115 V.A.C.: Controller: 72 Watts typical; Incubator: 640 Watts typical; Low Temperature Cart: 660 Watts typical

    At 230 V.A.C.: Controller: 72 Watts typical; Incubator: 640 Watts typical; Low Temperature Cart: 605 Watts typical

     

    Heat dissipated:

     

    Controller: 245 BTU/hr, 2840 maximum

    Incubator: 904 BTU/hr, 2840 maximum

    Low Temperature Cart: 2249 BTU/hr at 660 watts

    Low Temperature Cart: 2061 BTU/hr at 605 watts

     

    Controller Module dimensions:

     

    Width: 35.6 cm (14 in)

    Height: 91.4 cm (36 in)

    Depth: 61.7 cm (24.3 in)

    Weight: 57.2 kg (126 lbs)

     

    Incubator Module dimensions:

     

    Width: 49.6 cm (19.5 in)

    Height: 91.4 cm (36 in)

    Depth: 61.7 cm (24.3 in)

    Weight unloaded: 118.8 kg (262 lbs)

    Weight loaded: 133.8 kg (295 lbs)

     

    Low Temperature Module dimensions:

     

    Width: 80.9 cm (31.9 in)

    Height: 184.6 cm (72.7 in)

    Depth: 66.1 cm (26.0 in)

    Weight unloaded: 293.5 kg (647 lbs)

    Weight loaded: 313.9 kg (692 lbs)

     

    Benchtop cart dimensions:

     

    Width: 99.1 cm (39 in)

    Height: 85.1 cm (33.5 in)

    Depth: 63.5 cm (25 in)

    Weight: 100 kg (220 lbs)

     

    System operating and storage humidity range:

     

    Storage: 20-85% relative humidity, non-condensing

    Operating: 20-60% relative humidity, non-condensing

     

    System operating and storage altitude:

     

    6562 feet (2000 meters)

    Pioneering diagnostics